JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


Anesth Analg 1979; 58:30-32
© 1979 International Anesthesia Research Society
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nawaf, K.
Right arrow Articles by Stoelting, R. K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nawaf, K.
Right arrow Articles by Stoelting, R. K.

Nitrous Oxide Increases Enflurane Concentrations Delivered by Ethrane Vaporizers

Kays Nawaf, MD*, and Robert K. Stoelting, MD{dagger}

*Anesthesia Resident IV, Department of Anesthesia, Indiana University School of Medicine, Indianapolis, Indiana 46223. {dagger}Professor and Chairman, Department of Anesthesia, Indiana University School of Medicine, Indianapolis, Indiana 46223.

Abstract

Delivered enflurane concentrations from two calibrated Ethrane vaporizers were determined with total gas flows of 3, 5 and 8 L/min. Regardless of total gas flow the presence of 60% nitrous oxide increased enflurane concentrations by 20 to 40% above those concentrations present when only oxygen was flowing through the vaporizer. This nitrous oxide effect was present at all dial settings studied except the lowest engraved (0.25) concentration. Enflurane output at the 0.25% setting was 0.38% with or without nitrous oxide. Maximum changes in enflurane concentrations after adding nitrous oxide required about 5 minutes but the rapidity with which enflurane concentrations approached this maximum value were directly related to total gas flow. Similar effects of nitrous oxide on enflurane output from Cyprane Ethrane vaporizers were also measured.

The mechanism of increased vaporizer enflurane output in the presence of nitrous oxide is unknown but may reflect increased gas flow through the vaporizing chamber secondary to increase in gas density associated with nitrous oxide. A similar mechanism has been proposed to explain increased halothane concentrations delivered by Fluotec Mark 2 vaporizers in the presence of nitrous oxide.

Clinically, central nervous system stimulation and anesthetic overdose may occur from increased enflurane concentrations delivered when nitrous oxide is added to the gases flowing through the Ethrane vaporizer. The ability to deliver low enflurane concentrations is limited since the measured concentration at the lowest dial setting was nearly 0.4%.

Key Words: EQUIPMENT: vaporizers • ANESTHETICS, Volatile: enflurane • ANESTHETICS, Gases: nitrous oxide







Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins with the assistance of Stanford University Libraries' HighWire Press®. Copyright 2006 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press