JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


Anesth Analg 1979; 58:216-220
© 1979 International Anesthesia Research Society
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chalon, J.
Right arrow Articles by Turndorf, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chalon, J.
Right arrow Articles by Turndorf, H.

Humidification in a Modified Circle System

Jack Chalon, MD*, Clifford Goldman{dagger}, Mahnoosh Amirdivani, MD{ddagger}, Amy Rothblatt, BA{dagger}, Sivam Ramanathan, MD§, and Herman Turndorf, MD||

*Professor and Associate Director. Department of Anesthesiology, New York University Medical Center, New York, New York 10016 {dagger}Research Assistant. Department of Anesthesiology, New York University Medical Center, New York, New York 10016 {ddagger}Fellow. Department of Anesthesiology, New York University Medical Center, New York, New York 10016 §Assistant Professor. Department of Anesthesiology, New York University Medical Center, New York, New York 10016 ||Professor and Chairman. Department of Anesthesiology, New York University Medical Center, New York, New York 10016

Abstract

An improved circle system, capable of delivering high inspiratory humidity from the onset of anesthesia, was made by introducing a water vaporizer into the soda lime canister and tested on a model patient. Vaporization was facilitated by the exothermic reaction of neutralization of the lime. Humidity loss due to rain out in the absorber and inspiratory limb was minimized by constructional alterations of the absorber circuit assembly: (1) the conduit pipe which leads the moist absorber gases into the inspiratory limb was moved from its conventional outside position to the warm interior of the lime canister; (2) the moisture-laden fresh gases emerging from the vaporizer were delivered close to the patient by a small bore tube placed concentrically inside the inspiratory limb; and (3) both the fresh and inspired gases were thermally insulated by an outermost, moist, warm coaxial expiratory limb.

The humidity output of the circuit was high 1 minute after starting ventilation, averaging 19.6 ± 2, 19.3 ± 2.9, and 19 ± 1.1 mg of H2O/L as carbon dioxide output was increased by 100-ml increments from 100 to 300 ml/min (using fresh gas inflow of 5 L/min and a ventilatory minute volume of 6 L/min). After a 2- to 3-hour period of stabilization, humidity reached 22.8 ± 1.5, 24 ± 0.9, and 28 ± 2.4 mg of H2O/L at the carbon dioxide flows quoted above. Vaporizer patent held by Albert Einstein College of Medicine, Bronx, NY (circuit patent pending).

Key Words: HUMIDIFICATION: anesthetic gases • EQUIPMENT: humidification







Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins with the assistance of Stanford University Libraries' HighWire Press®. Copyright 2006 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press