Received from the Anesthesia Services of Massachusetts General Hospital and the Department of Anesthesia, Harvard Medical School, Boston, Massachusetts; and the Department of Chemistry, Oberlin College, Oberlin, Ohio.
Abstract
A randomized, blind study examined the effect of nail polish color on measurement of oxygen saturation by pulse oximetry. Fourteen adult volunteers had blue, green, purple, black, and red nail polish applied to their finger nails. A strip-chart recording of oxygen saturation (Nellcor N100) was made in room air and later interpreted in a blinded fashion. The absorption spectra of the five polishes were determined by spectrophotometry. The spectra of nine other nail polishes and three intravenous dyes also were examined. Black, blue, and green nail polish significantly lowered oximeter readings of oxygen saturation. Blue and green produced greater decreases than purple and red; black produced an intermediate decrease. Some but not all nail polishes absorbed light at the wavelengths used by the pulse oximeter (660 nm and 940 nm). The degree of artifactual desaturation correlated best with the difference between absorbance at 660 nm and absorbance at 940 nm (r = 0.95). Spectrophotometric absorbance data suggest that other colors may interfere with pulse oximetry. On the basis of Spectrophotometric data, brown-red nail polish was predicted to interfere with oximetry; subsequent pulse oximetry measurements confirmed the prediction. Nail polish should be removed routinely before pulse oximetry monitoring.
Key Words: OXYGEN, MEASUREMENT—oximetry MONITORING—pulse oximetry MEASUREMENT TECHNIQUES—pulse oximetry
This article has been cited by other articles:
![]() |
L. Zhang, S. Bhattacharya, C. J. Cote, and R. R. Anderson Alexandrite 755 nm Laser-Induced Pulse Oximetry Dysfunction Anesth. Analg., January 1, 2009; 108(1): 386 - 387. [Full Text] [PDF] |
||||
![]() |
J. Hinkelbein and H. V. Genzwuerker Hypercapnia as the influencing factor of pulse oximetry Eur. Respir. J., December 1, 2008; 32(6): 1670 - 1670. [Full Text] [PDF] |
||||
![]() |
P. B. Batchelder and D. M. Raley Maximizing the Laboratory Setting for Testing Devices and Understanding Statistical Output in Pulse Oximetry Anesth. Analg., December 1, 2007; 105(6S_Suppl): S85 - S94. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Gainet, I. Thaon, V. Westeel, H. Chaudemanche, A. G. Venier, A. Dubiez, J. J. Laplante, and J-C. Dalphin Twelve-year longitudinal study of respiratory status in dairy farmers Eur. Respir. J., July 1, 2007; 30(1): 97 - 103. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Paige and P. E. Bickler Falsely Low Pulse Oximetry Values in Patients Receiving Docetaxel (Taxotere(R)) Anesth. Analg., August 1, 2004; 99(2): 622 - 623. [Full Text] [PDF] |
||||
![]() |
M. M. Chan, M. M. Chan, and E. D. Chan What Is the Effect of Fingernail Polish on Pulse Oximetry? Chest, June 1, 2003; 123(6): 2163 - 2164. [Full Text] [PDF] |
||||
![]() |
M. J. Grap Pulse Oximetry Crit. Care Nurse, June 1, 2002; 22(3): 69 - 74. [Full Text] [PDF] |
||||
![]() |
D. A. Benaron, W. E. Benitz, R. L. Ariagno, and D. K. Stevenson Noninvasive Methods for Estimating In Vivo Oxygenation Clinical Pediatrics, May 1, 1992; 31(5): 258 - 273. [Abstract] [PDF] |
||||
![]() |
F. J. Curley and N. A. Smyrnios Routine Monitoring of Critically I11 Patients J Intensive Care Med, July 1, 1990; 5(4): 153 - 174. [Abstract] [PDF] |
||||
|